Magnetic Field Device (MFD)

Last Updated: February 3rd, 2002


The MFD magnetic field was calibrated on January 29th, 2002 at the University of North Florida under the supervision of Professor Tom Pekarek.  The nominal field strength in the center of the experiment area was 300 Gauss.  The field strength was reasonably consistent from the bottom of the lower petri dish to the top of the upper petri dish in the experiment area.
The magnetic field was calibrated at UNF on January 29th, 2002 using an Applied Magnetics Laboratory Gauss meter
UNF Professor Tom Pekarek assiting Amanda with the calibration
Calibrating the field at the bottom of the experiment area
Calibrating the field at the top of the experiment area
Each area of the experiment field was calibrated
The shape of the magnetic field was observed using a Nakamura Magnetic Field Observation Box

Calibration of the Magnetic Field

Calibrating the field at the top of the experiment area
Detailed Results
Magnetic Field Calibration

Measurement of Magnetic Field in Gauss units
Side A

Bottom of Lower Petri Dish
Between Two Petri Dishes
Top of Upper Petri Dish

 

-242   -289   -290   -230
____________________

+231  +292  +291  +228



 

-249   -303   -297  -241 ____________________

+245   +303   +302   +242

-213   -251   -244   -197
____________________

+208   +254   +246   +210

Side B

Description of the Magnetic Field Device (MFD)


 
 